| Input Voltage |
90 ~ 264Vac 127 ~ 370Vdc |
| Input Frequency |
47 ~ 63Hz |
| Power Factor |
PF>0.93/230Vac PF>0.98/115Vac at full load |
| Input Current |
2A/115Vac 1.1A/230Vac |
| Inrush Current |
35A/115Vac 70A/230Vac, cold start |
| Leakage Current |
Earth leakage current <300uA/ 264Vac, Patient leakage current <100uA/264Vac |
| Line Regulation |
±0.5% |
| Load Regulation |
±1.0% |
| Set Up, Rise Time |
1200ms, 30ms/230Vac 2500ms, 30ms/115Vac at full load |
| Hold Up Time |
16ms/230Vac/115Vac at full load |
| Over Load Protection |
105 ~ 135% rated output power |
|
Hiccup mode, recovers automatically after fault condition is removed |
| Over Voltage Protection |
Shut down o/p voltage, re-power on to recover |
| Over Temperature Protection |
110°C(5V),105°C(12V,15V,24V,48V) (TSW1 : detect on heatsink of power transistor) |
|
95°C(5V),90°C(12V,15V,24V,48V) (TSW2 : detect on heatsink of power transistor) |
|
(TSW1)Shut down o/p voltage, recovers automatically after temperature goes down |
|
(TSW2)Shut down o/p voltage, re-power on to recover |
| PS-ON Input Signal (Optional) |
Power on: PS-ON = Hi or >2V ; Power off: PS-ON = Low or <0.5V |
| Power Good/Power Fail |
500ms>PG>10ms PF>1ms |
| Remote Sense |
5 ~ 15V |
| Working Temperature |
-20°C ~ +70°C (Refer to output load derating curve) |
| Working Humidity |
20 ~ 90% RH non-condensing |
| Storage Temperature, Humidity |
-40°C ~ +85°C, 10 ~ 95% RH |
| Temperature Coefficient |
±0.03%/°C (0 ~ 50°C) |
| Vibration |
10 ~ 500Hz, 2G 10min./1cycle, 60min. each along X, Y, Z axes |
| Withstand Voltage |
I/P-O/P:4KVAC I/P-FG:1.5KVAC O/P-FG:0.5KVAC |
| Isolation Resistance |
I/P-O/P, I/P-FG, O/P-FG:100M Ohms / 500VDC/ 25/°C 70% RH |
| Safety Standards |
UL60601-1, TUV EN60601-1 approved |
| EMC |
Compliance to EN55011 (CISPR11), EN55022 (CISPR22) Class B; EN61000-3-2,-3; ENV50204; |
|
EN61000-4-2,3,4,5,6,8,11; EN55024, EN60601-1-2, EN61204-3, medical level, Criteria A |
| MTBF |
230.5Khrs min. MIL-HDBK-217F (25°C) |
| Dimensions |
127 x 76.2 x 34.6mm |